Answer
Jul 15, 2014 - 12:36 AM
The Bump Tests are no longer available. You are able to Bump Test your unit by your breath following the below instructions from the manufacturer. You can also find these directions in the product manual under section 3.2 by following this URL:
http://www.analox.net/product-docs/ma...
Bump test your EII CO with your breath
1. Power the EII CO up and set the display to zero
2. Take a deep breath and hold it for 5 to 10 seconds
3. With the dome adaptor removed, exhale a slow and even breath into the dome
aperture
4. Ensure that the displayed reading rises to 1 to 2ppm (smokers may see a higher
sensor response)
5. If breath testing fails to raise the display reading sufficiently, then a traditional bump
test should be performed as per section 3.2.1. Please see Section 5 for further
troubleshooting information.
http://www.analox.net/product-docs/ma...
Bump test your EII CO with your breath
1. Power the EII CO up and set the display to zero
2. Take a deep breath and hold it for 5 to 10 seconds
3. With the dome adaptor removed, exhale a slow and even breath into the dome
aperture
4. Ensure that the displayed reading rises to 1 to 2ppm (smokers may see a higher
sensor response)
5. If breath testing fails to raise the display reading sufficiently, then a traditional bump
test should be performed as per section 3.2.1. Please see Section 5 for further
troubleshooting information.